Droplet Analyser compulsory requirements:
- measures the size (diameter) of freely moving water droplets in real-time
- uses optical measurement principle
- does not interfere the droplet motion or affect the droplet size (beyond normal evaporation)
- has immediate measurement range covering droplet diameters from 3 μm to 100 μm
- has size resolution and accuracy for water droplets better than +/- 0.5 μm for the whole measurement range (95% confidence interval)
- has NIST traceable calibration for water droplets
- is applicable to in-situ use outside conventional laboratory environment
- includes all necessary signal processing hardware and PC software running on Windows 10 for instrument control, data acquisition and data presentation/analysis (excluding PC-computer)
- has an expected lifetime of at least 10 000 hours
- includes extended warranty with minimum duration of 24 months
- has total price 70 000 € or less (excluding VAT)
• Tenders not fulfilling all the compulsory requirements will not be taken into account.
• The tender should include brief explanation showing that the spectrometer fulfils each of the compulsory requirements e.g. by pointing out the facts from the instrument/device data sheet, test reports or application reports etc.